Campbell CR5000 Especificações Página 115

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 238
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 114
Section 6. Data Table Declarations and Output Processing Instructions
6-21
Point Source Action Bin 1
(level=1)
Bin 2
(level=1.5)
Bin 3
(level=3)
1 0.5 First value, no
counts
00 0
2 1.2 Add one count to
first bin, the
signal crossed 1
10 0
3 1.4 No levels crossed,
no counts
10 0
4 0.3 Crossed a level
but was falling
edge, no counts
10 0
5 3.3 Add one count to
first, second, and
third bins, the
signal crossed 1,
1.5 and 3.
21 1
The second dimension, when greater than 1, is determined by the value of the
element in the source array following the element checked for the crossing. It
is the value of this variable at the time the crossings are detected that
determines the second dimension.
0
0.5
1
1.5
2
2.5
3
3.5
0123456
Crossing Levels
Crossing Source
2nd Dim Boundary
2ndDim Source
FIGURE 6.4-2. Crossing Data with Second Dimension Value
Vista de página 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 237 238

Comentários a estes Manuais

Sem comentários