Appendix C. Interfacing with HMP155A RS-485 Output
The public variables for temperature and relative humidity can be viewed in the
Numeric Displaymode as shown below.
For troubleshooting purposes, the serial data buffer in the datalogger can be
viewed using the ‘W’ terminal command. This is done by connecting to the
datalogger from the Connectbutton of LoggerNet or PC400W. From the
Connect screen, select Tools|Terminal Emulator. Click the Open Terminal
button, and press the enter key to get the CR1000prompt. Type ‘W’ for the
Serial Comms Sniffer. Select ‘4’ for ComSDC7, and ‘Y’ for ASCII. Raw
serial data received by the buffer is displayed on the screen as shown below.
C-9