Appendix A. Discussion of TDR Probe Offset and a Simple Laboratory Method for Calculation
4. Enter Terminal Mode using Options/Terminal Emulator.
5.
Press Enter until > appears (the line commands are not case sensitive)
6.
Type GVAR then Enter.
7.
It is recommended that step 6 be repeated several times and that the
average values of Start and End be used for following calculations.
GVAR returns the uncorrected
Start and End. These values must be
converted to distance from index values. This is done as follows:
ngthWaveformLe
1datapoints
end
:endngthWaveformLe
1datapoints
start
:start
distancedistance
=
=
Equations [A3] and [A4] are then used to calculate probe offset.
A.3.2 An Example Using CS605
Measured TDR probe rod length: 3.0:L
=
m.
Measure temperature of water in column C4.24:T °
=
.
Determine actual dielectric permittivity of water using following equation
or values in TABLE A-1.
() () () ()
+=ε
3
8
2
5
3
25T
10
8.225T
10
19.125T
10
5791.41
54.78:T
()
76.78T =ε
(
)
TL:La ε=
m 66 .2La =
Waveform parameters for PC-TDR
0.1:Vp251:datapointsm 5:thWindowLeng
=
=
=
m 0offset Probe m 0.3length Probe
=
=
Start and end index values from terminal emulator mode of PC-TDR as
described above
87.169:end44.32:start
indexindex
=
=
converting waveform index to apparent distance
thWindowLeng
1datapoints
start
:start
index
distance
=
A-4