Number 4250-2000
Revision 1.0
Date MAY 1996
Page 2 of 7
Ensure that all datalogger servicing is documented and archived in accordance with the
procedures described in this TI.
2.2 INSTRUMENT TECHNICIAN
The instrument technician shall:
Perform and document all calibration checks.
Coordinate with Campbell Scientific for return and recalibration of Campbell 21X
dataloggers that fail to operate within factory specifications.
Prepare purchase orders for factory servicing and recalibration of Campbell 21X
dataloggers.
Replace the Campbell 21X internal battery as required.
Archive all datalogger servicing records.
2.3 DATA COORDINATOR
The data coordinator shall:
Inform the instrument technician when a 21X is being removed from the field.
Provide the instrument technician with a description of the field problems observed with
the 21X.
3.0 REQUIRED EQUIPMENT AND MATERIALS
Specific instrumentation, tools, equipment, and materials required to service the Campbell
21X datalogger and to verify the datalogger calibration are:
Calibrated voltage source - Datel Model DVC-350A or equivalent
Campbell Scientific datalogger communications software (SMCOM)
Campbell Scientific SC532 Peripheral Interface Module
ARS Campbell 21X datalogger test program (21X_TEST.DLD)
Digital voltmeter (4 1/2 digits)
Waveform generator - Wavetek Model 185 or equivalent
Frequency counter - Tenma Model 72-375 or equivalent
Campbell Scientific, Inc.
21X Micrologger Operator's Manual
and
21X Prompt Sheet
Laboratory reference thermometer (
°
C)