Model HFP01SC Self-Calibrating Soil Heat Flux Plate
;Factory calibration in W/(m^2 mV) = 1000/sensitivity.
;
4: Z=F (P30)
1: 1 F ; <- Enter the unique calibration here
2: 0 Exponent of 10
3: 3 Z Loc [ cal ]
5: End (P95)
;Use data from the larger measurement range.
;
6: If (X<=>F) (P89)
1: 2 X Loc [ shf_mV ]
2: 4 <
3: -99990 F
4: 30 Then Do
7: Z=X (P31)
1: 8 X Loc [ shf_mV_a ]
2: 2 Z Loc [ shf_mV ]
8: End (P95)
;Apply custom calibration to the raw soil heat flux measurement.
;
9: Z=X*Y (P36)
1: 2 X Loc [ shf_mV ]
2: 3 Y Loc [ cal ]
3: 1 Z Loc [ shf ]
;Output data.
;
10: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 20 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
11: Real Time (P77)
1: 0110 Day,Hour/Minute (midnight = 0000)
12: Resolution (P78)
1: 1 High Resolution
;Do not include that calibration data in the soil heat flux.
;
13: If Flag/Port (P91)
1: 118 Do if Flag 18 is High
2: 19 Set Intermed. Proc. Disable Flag High (Flag 9)
14: Average (P71)
1: 1 Reps
2: 1 Loc [ shf ]
15: Do (P86)
1: 29 Set Intermed. Proc. Disable Flag Low (Flag 9)
16: Sample (P70)
1: 1 Reps
2: 3 Loc [ cal ]
;Add other processing here.
;Call calibration routine.
;
17: Do (P86)
1: 8 Call Subroutine 8
14