Resumo do Conteúdo

Página 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL CNR4 Net Radiometer Revision: 2/15 Copyright © 2000- 2015 Campbell Scientific, Inc.

Página 2

Table of Contents iv

Página 3 - Limited Warranty

CNR4 Net Radiometer 1. Introduction The CNR4 is a research-grade net radiometer that measures the energy balance between incoming and outgoing radia

Página 4 - Assistance

CNR4 Net Radiometer 4. Quickstart Please review Section 7, Operation, for wiring and CRBasic programming. Appendix D, CNF4 Heater/Ventilator, provid

Página 5 - Precautions

CNR4 Net Radiometer 2. Attach the 26120 mounting bracket to the vertical mounting pipe, or CM200-series crossarm using the provided U-bolt (see FIGUR

Página 6

CNR4 Net Radiometer 5. Tighten the four screws on top of the mounting bracket to properly secure the support arm so that it does not rotate (see FIGU

Página 7

CNR4 Net Radiometer 3. When Short Cut opens, select New Program. 4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for most

Página 8

CNR4 Net Radiometer 5. Under the Available Sensors and Devices list, select Sensors | Meteorological | Solar Radiation folder. Select CNR4 Net Radia

Página 9

CNR4 Net Radiometer 7. After selecting the sensitivity values, click at the left of the screen on Wiring Diagram to see how the sensor is to be wired

Página 10 - Table of Contents

CNR4 Net Radiometer Two temperature sensors, a thermistor and a Pt-100, are integrated with the CNR4 body. The temperature sensor is used to provide

Página 11 - CNR4 Net Radiometer

CNR4 Net Radiometer • Separate outputs of short-wave and long-wave infrared radiation for better accuracy and more thorough quality assurance • Sola

Página 13

CNR4 Net Radiometer 6.1 CNR4 Specifications Sensor sensitivities: Four probes with unique sensitivity values. Please refer to the calibration sheets

Página 14 - 4.3 Short Cut Programming

CNR4 Net Radiometer Directional error*: < 20 W/m2 at angle up to 80° with 1000 W/m2 Zero offset due to 0 to -200 W/m2 IR net irradiance*: < 15

Página 15

CNR4 Net Radiometer Field of view Upper: Lower: 180 degrees 150 degrees Net-irradiance: –250 to +250 W/m2 Non-stability: < 1% (sensitivity change

Página 16

CNR4 Net Radiometer 7.1 Using the CNR4 in the Four Separate Components Mode In the four separate components mode configuration (measuring two short-w

Página 17 - 5. Overview

CNR4 Net Radiometer In this equation, C is the sensitivity of the sensor. T is in Kelvin, and not in Celsius or Fahrenheit. The downward-facing pyrg

Página 18 - 6. Specifications

CNR4 Net Radiometer TABLE 7-1. Resistance values versus CNR4’s thermistor temperature in °C Temperature [°C] Resistance [Ω] Temperature [°C] Resistan

Página 19

CNR4 Net Radiometer TABLE 7–2. Resistance values versus CNR4’s Pt–100 temperature in °C Temperature [°C] Resistance [Ω] Temperature [°C] Resistance [

Página 20 - 6.1 CNR4 Specifications

CNR4 Net Radiometer site, the calculation of albedo can also serve as a tool for quality control of your measured data at a specific site. 7.1.5 Calc

Página 21 - to 0 to -200 W/m

CNR4 Net Radiometer 7.1.7 Calculation of Net (Total) Radiation In the four separate components mode, net radiation, Rn, can be calculated using the i

Página 22 - 7. Operation

CNR4 Net Radiometer FIGURE 7-2. The marks on the end of the CNR4: S for SOLAR cable, and T for TEMP cable The measurement details for Pt-100 sensor,

Página 23

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Página 24

CNR4 Net Radiometer FIGURE 7-4. Labels on the pigtail end of the TEMP cable. TABLE 7-3. Datalogger Connections for Differential Measurement Functi

Página 25

CNR4 Net Radiometer TABLE 7-4. Datalogger Connections for Single-Ended Measurement Function Wire Color CR6, CR1000, CR3000, CR5000 Pyranometer Up S

Página 26 - 7.1.4 Calculation of Albedo

CNR4 Net Radiometer The CNR4 outputs four voltages that typically range from 0 to 15 mV for the pyranometers, and ± 5 mV for the pyrgeometers. A diff

Página 27

CNR4 Net Radiometer 8.1 Testing the Pyranometer As a first test, check the sensor impedance. It should have a nominal value as indicated in the spec

Página 28 - 7.2 Wiring

CNR4 Net Radiometer 8.4 Testing the Pt-100 Using a multimeter, measure the resistance between the two opposite wires of the Pt-100 (gray-yellow, gray

Página 29

CNR4 Net Radiometer FIGURE 9-1. Replacing the drying cartridge 9.4 Replacement Parts The following is the list of replacement parts for the CNR4 an

Página 30

CNR4 Net Radiometer 26

Página 31 - 7.3 Datalogger Programming

Appendix A. Importing Short Cut Code Into CRBasic Editor This tutorial shows: • How to import a Short Cut program into a program editor for additio

Página 32 - 8. Troubleshooting

Appendix A. Importing Short Cut Code Into CRBasic Editor mark) begins each line. This character instructs the datalogger compiler to ignore the line

Página 33 - 8.3 Testing the Thermistor

Appendix B. Example Programs B.1 CR1000 Program Using Differential Measurements This example requires four differential channels to measure the fou

Página 34 - 9.2 Recalibration

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Página 35 - 9.4 Replacement Parts

Appendix B. Example Programs ' '8H '8L CNR4 thermistor signal (white) 'gnd CNR4 thermistor signal reference (black)

Página 36

Appendix B. Example Programs Sample (1,logger_temp,FP2) Average (4,cnr4(1),IEEE4,False) Average (1,cnr4_T_C,IEEE4,False) Average (1,cnr4_T_K,

Página 37 - Into CRBasic Editor

Appendix B. Example Programs B.2 CR3000 Program Using Differential Measurements This example requires four differential channels to measure the four

Página 38

Appendix B. Example Programs 'VOLTAGE EXCITATION ' 'VX1 CNR4 thermistor voltage excitation (red) ' 'CNR4 sensor Public

Página 39 - Measurements

Appendix B. Example Programs Average (1,Rs_net,IEEE4,False) Average (1,Rl_net,IEEE4,False) Average (1,albedo,IEEE4,False) Average (1,Rn,IEEE4

Página 40

Appendix B. Example Programs B.3 CR5000 Program Using Differential Measurements This example requires four differential channels to measure the four

Página 41

Appendix B. Example Programs '4H CNR4 Pyrgeometer Lower signal (brown) '4L CNR4 Pyrgeometer Lower signal reference (green) 'gn

Página 42

Appendix B. Example Programs Minimum (1,batt_volt,FP2,0,False) Sample (1,logger_temp,FP2) Average (4,cnr4(1),IEEE4,False) Average (1,cnr4

Página 43

Appendix B. Example Programs B.4 CR3000 Program for Measuring Pt-100 Temperature Sensor This example measures the Pt-100 sensor for the body tempera

Página 44

Appendix B. Example Programs 'CR3000 Series Datalogger ' 'CNR4 program 'This program measures CNR4 four-component net radiometer

Página 45

Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO

Página 46

Appendix B. Example Programs Units cnr4_T_K = K Units long_up_corr = W/m^2 Units long_dn_corr = W/m^2 Units Rs_net = W/m^2 Units Rl_net = W/m^2 Units

Página 47

Appendix B. Example Programs PRT (cnr4_T_C,1,Rs_R0,1,0) 'Convert CNR4 temperature to Kelvin cnr4_

Página 48 - Temperature Sensor

Appendix B. Example Programs B-14

Página 49

Appendix C. CNR4 Performance and Measurements under Different Conditions TABLE C-1 shows what one might typically expect to measure under different

Página 50

Appendix C. CNR4 Performance and Measurements under Different Conditions FIGURE C-1. Different measurement conditions and signals FIGURE C-2. Par

Página 51

Appendix C. CNR4 Performance and Measurements under Different Conditions FIGURE C-3. Clear day for the downward facing pyrgeometer It is assumed th

Página 52

Appendix C. CNR4 Performance and Measurements under Different Conditions 4/181067.5CG3upper EeTemperaturSky ⋅=− (C-1) 4/181067.5⋅=−CG

Página 53 - Conditions

Appendix D. CNF4 Heater/Ventilator Whenever the heater is used, the heating may cause errors in the measurement of the sensor temperature. Under mos

Página 54 - Upper pyrgeometer

Appendix D. CNF4 Heater/Ventilator 10 watt power available? Not available DO NOT HEAT Available Consider options below Clock and relay available?

Página 55

Appendix D. CNF4 Heater/Ventilator D.2 Attaching the Optional CNF4 Heater/Ventilator Unit to CNR4 1. The CNF4 heater/ventilator unit comes with the

Página 57 - D.1 General Information

Appendix D. CNF4 Heater/Ventilator 2. Attach the heater/ventilator unit unto the bottom of the CNR4 sensor, using the three pan-head screws and wash

Página 58

Appendix D. CNF4 Heater/Ventilator 3. Make sure the cables are cleared from the edges of the CNF4, as shown in FIGURE D-3, and place the white solar

Página 59 - Unit to CNR4

Appendix D. CNF4 Heater/Ventilator FIGURE D-5. Attaching the solar shield to CNF4 using four flat-head screws 4. Once the CNF4 heater/ventilator u

Página 60

Appendix D. CNF4 Heater/Ventilator D.3 Wiring The following table shows the recommended datalogger wiring for using the CNR4 sensor with the CNF4 he

Página 61

Appendix D. CNF4 Heater/Ventilator D.4 CR3000 Datalogger Program with Heater/Ventilator Control This example measures the four radiation outputs, th

Página 62

Appendix D. CNF4 Heater/Ventilator ' CNF4 heater - (yellow) ' 'gnd ventilator & heater shield (clear) Pip

Página 63 - D.3 Wiring

Appendix D. CNF4 Heater/Ventilator Average (1,long_dn_corr,IEEE4,False) Average (1,Rs_net,IEEE4,False) Average (1,Rl_net,IEEE4,False) Average

Página 64 - Heater/Ventilator Control

Appendix D. CNF4 Heater/Ventilator D.5 CNF4 Heater/Ventilator Maintenance D.5.1 Testing the Heater The optional CNF4 consists of a heater and a ven

Página 65

Appendix D. CNF4 Heater/Ventilator D-12

Página 67 - D.5.1 Testing the Heater

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Página 68

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Página 69

Table of Contents 9. Maintenance and Recalibration ... 24 9.1 Cleaning Windows and Domes ...

Página 70

Table of Contents D-4. CNF4 solar shield and four flat-head screws ......... D-5 D-5. Attaching the solar shield to CNF4 us

Comentários a estes Manuais