Campbell CS455 Manual Página 22

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 21
CS450/CS455 Pressure Transducer
5.2 RS-232
TABLE 5-2. RS-232 Wiring
Color
CS450/CS455
Function
Connection
RS-232 9-pin
Red +12vdc Power Source
Black Power Ground Power Ground
White RS-232 Tx (Output) Transmit Pin 2 Rx (Input)
Blue RS-232 Rx (Input) Receive Pin 3 Tx (Output)
Yellow Digital Ground Ground Pin 5 GND
Clear Shield GND Ground
6. Programming
6.1 Using SCWIN
SCWIN is the easiest and typically the preferred method for programming the
datalogger. SCWIN generates a wiring diagram that shows how to connect the
CS450/SC455 to your Campbell Scientific datalogger.
The sections that immediately follow are for CRBasic and Edlog.
NOTE
Keyboard/Display users. SCWIN users can jump ahead to the Maintenance
section.
6.2 Using CRBasic or Edlog
6.2.1 CRBasic Programming
Dataloggers that use CRBasic include our CR200-series, CR800, CR850,
CR1000, CR3000, and CR5000. These dataloggers use the SDI12Recorder to
read the CS450/CS455. A multiplier of 1.0 and an offset of 0.0 yield water
level in psig and temperature in degrees C.
The SDI12Recorder instruction has the following form:
SDI12Recorder(Destination, Output String, Multiplier, Offset)