CS616 and CS625 Water Content Reflectometers
;{CR10X}
;
*Table 1 Program
01: 3600 Execution Interval (seconds)
1: Batt Voltage (P10)
1: 7 Loc [ batt ]
2: CS616 Water Content Reflectometer (P138)
1: 3 Reps
2: 1 SE Channel
3: 6 C6 is first of sequential Control Ports used
4: 1 Loc [ period_1 ]
5: 1.0 Mult
6: 0.0 Offset
3: Polynomial (P55)
1: 3 Reps
2: 1 X Loc [ period_1 ]
3: 4 F(X) Loc [ vwc_1 ]
4: -0.0663 C0
5: -0.0063 C1
6: 0.0007 C2
7: 0.0 C3
8: 0.0 C4
9: 0.0 C5
;
4: If time is (P92)
1: 0000 Minutes (Seconds --) into a
2: 240 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
5: Average (P71)
1: 4 Reps
2: 4 Loc [ vwc_1 ]
*Table 2 Program
02: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
End Program
-Input Locations-
1 period_1 5 1 1
2 period_2 9 1 1
3 period_3 17 1 1
4 vwc_1 5 1 1
5 vwc_2 9 1 1
6 vwc_3 17 1 1
7 batt 1 0 1
6.4.5.4 CS616 Sample Edlog Program 4
Measure soil temperature with 107 probe and CS616 output period every four
hours. Correct CS616 output period for temperature and write result to
datalogger final storage.
19