Campbell CR510 Manual do Operador Página 26

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 264
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 25
CR510 OVERVIEW
OV-14
OV5.2 EDITING AN EXISTING PROGRAM
When editing an existing program in the CR510,
entering a new instruction inserts the
instruction; entering a new parameter replaces
the previous value.
To insert an instruction, enter the program table
and advance to the position where the
instruction is to be inserted (i.e., P in the data
portion of the display) key in the instruction
number, and then key A. The new instruction
will be inserted at that point in the table,
advance through and enter the parameters.
The instruction that was at that point and all
instructions following it will be pushed down to
follow the inserted instruction.
An instruction is deleted by advancing to the
instruction number (P in display) and keying #D
(Table 4.2-1).
To change the value entered for a parameter,
advance to the parameter and key in the correct
value then press A. Note that the new value is
not entered until A is keyed.
SAMPLE PROGRAM 2
Instruction # Parameter
(Loc:Entry) (
Par#:Entry) Description
1 Enter Program Table 1
01:60 60 second (1 minute) execution interval
Key
#
D
until 01:P00 Erase previous Program before
is displayed continuing.
01:P11 Measure reference temperature
01:1 Store temp in Location 1
02:5
03:3
04:1
05:1.0
06:0.0
02:P92 If Time instruction
01:0 0 minutes into the interval
02:60 60 minute interval
03:10 Set Output Flag 0
The CR510 is programmed to measure the datalogger internal temperature every sixty seconds.
The If Time instruction sets the Output Flag high at the beginning of every hour. Next, the
Output Instructions for time and average are added.
03:P77 Output Time instruction
01:110 Store Julian day, hour, and minute
04:P71 Average instruction
01:1 one repetition
02:1 Location 1 - source of temps. to be averaged
05:P92 If Time instruction
01:0 0 minutes into the interval
02:1440 1440 minute interval (24 hrs.)
03:10 Set Output Flag 0
06: P77 Output Time instruction
01:100 Store Julian day
Vista de página 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 263 264

Comentários a estes Manuais

Sem comentários