Campbell AM16 Manual Página 33

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 32
AM16/32B Relay Analog Multiplexer
*1 Table 1 Programs
1: 60 Sec. Execution Interval
;ACTIVATES MULTIPLEXER
1: Do (P86)
1: 41 Set high Port 1
;BEGINS STRAIN GAGE MEASUREMENT LOOP
2: Beginning of Loop (P87)
1: 0 Delay
2: 10 Loop Count
;CLOCK PULSE
3: Do (P86)
1: 72 Pulse Port 2
;DELAY
4: Excitation with Delay (P22)
1: 1 EX Chan
2: 0 ) Delay w/EX (units=.01sec
3: 1 Delay after EX (units=.01sec)
4: 0 mV Excitation
;FULL BRIDGE MEASUREMENT INSTRUCTION
5: Full Bridge (P6)
1: 1 Rep
2: 3 50 mV slow Range
3: 1 IN Chan
4: 1 Excite all reps w/Enchain 1
5: 5000 mV Excitation
6: 1-- Loc [:STRAIN #1]
7: 1 ult M
8: 0 Offset
;END OF STRAIN GAGE MEASUREMENT LOOP
6: End (P95)
;BEGINNING OF POTENTIOMETER MEASUREMENT LOOP
7: Beginning of Loop (P87)
1: 0 Delay
2: 6 Loop Count
8: Step Loop Index (Extended) (P90)
1: 2 Step
;CLOCK PULSE
9: Do (P86)
1: 72 Pulse Port 2
;DELAY
10: Excitation with Delay (P22)
1: 1 EX Chan
2: 0 Delay w/EX (units=.01sec)
3: 1 Delay after EX (units=.01sec)
4: 0 mV Excitation
23