Campbell HMP45C Guia do Utilizador Página 17

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 16
User Guide
9
CR10(X) Program using Single-Ended Measurement Instructions Using a PSW12 or SW12V on
Datalogger
;Turn the HMP45C on.
;
01: Do (P86)
1: 41 Set Port 1 High ;Jumper wire from PSW12 control to C1
;Yellow wire (C1) if using PSW12 device
;For CR23X or CR5000 use 49 for SW12V internal
;control port
;Pause 150 mSec before making measurements so the
;probe can stabilize on true readings.
;
02: Excitation with Delay (P22)
1: 1 Ex Channel
2: 0 Delay W/Ex (units = 0.01 sec)
3: 15 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation
;Measure the HMP45C temperature.
;
03: Volt (SE) (P1)
1: 1 Reps
2: 5 2500 mV Slow Range ;CR510, CR500 (2500mv); CR23X (1000 mV);
21X, CR7 (5000 mV)
3: 3 SE Channel
4: 1 Loc [ T_C ]
5: .1 Mult ;See Table 4 for alternative multipliers
6: -40 Offset ;See Table 4 for alternative offsets
;Measure the HMP45C relative humidity.
;
04: Volt (SE) (P1)
1: 1 Reps
2: 5 2500 mV Slow Range ;CR510, CR500 (2500 mV); CR23X (1000 mV);
21X, CR7 (5000 mV)
3: 4 SE Channel
4: 2 Loc [ RH_pct ]
5: .1 Mult ;See Table 5 for alternative multipliers
6: 0 Offset
;Turn the HMP45C off.
;
05: Do (P86)
1: 51 Set Port 1 Low
Vista de página 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 27 28

Comentários a estes Manuais

Sem comentários