November 2005, ARM-TR-069
B-3
SURFACE TEMPERATURE AND HUMIDITY
REFERENCE SYSTEM
SENSOR VERIFICATION
Tested by:______________________ Date: ________/_________/__________
Started: __________________GMT Ended:__________________GMT
1. Vaisala HMP-45D Temperature and RH
(The temperature readings, once stabilized should match within +/- 0.5°C)
(The RH readings, once stabilized should match within +/- 3% RH for 0 – 90% RH and
+/- 4% for 90 – 100% RH)
V1 HMP-45D Serial Number:______________
Temperature:
HMI31 Reading:___________V1 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________V1 HMP-45D Reading:_____________ P/F
V2 HMP-45D Serial Number_______________
Temperature:
HMI31 Reading:___________V2 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________V2 HMP-45D Reading:_____________ P/F
V3 HMP-45D Serial Number________________
Temperature:
HMI31 Reading:___________V3 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________V3 HMP-45D Reading:_____________ P/F