Campbell NL100 Manual do Utilizador Página 3

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 44
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 2
November 2005, ARM-TR-069
1
Surface Temperature Humidity Reference System
1. Purpose
The Surface Temperature and Humidity Reference (SURTHREF) System is intended to provide
accurate reference values of ambient temperature and relative humidity (RH) for comparison with
radiosonde pre-launch values.
2. References
ECR-00319 (Modify balloon Borne sounding system (BBSS) surface reference system at the
Southern Great Plains [SGP] site)
ECO-00319 (Modify the Chilled Mirror/BBSS surface reference system at the SGP site)
3. Components
The components of the SURTHREF system are as follows:
Meteorological Instrument Shelter. The shelter is a standard NWS “Stevenson screen”
modified to allow for operator access on two sides. The shelter dimensions are 20 inches
by 30 inches by 34 inches.
Instrument Shelter stand.
Aspirated Chamber. The chamber is fabricated from white polypropylene plastics. Its
dimensions are 11 inches by 13 inches by 18 inches (volume = 1.49 feet
3
). One end is
open and covered with a removable plate containing a 7-inch by 7.5-inch access port.
Two muffin fans, each rated at 115 cfm are mounted at the far end. The fans produce an
estimated face flow of 10.49 fps or 3.2 m/s. The chamber has six sensor ports on its top
surface to accommodate standard temperature and RH probes.
Sonde positioning platform. A polypropylene platform is mounted approximately 1 inch
above the bottom of the aspirated chamber. The platform is equipped with moveable
guides for positioning radiosondes. A user-controlled switch is mounted on the side of
the Aspirated Chamber to indicate when a radiosonde is placed in or removed from the
chamber.
Temperature and RH probes.
o Vaisala HMP45D probes – four each, one spare.
o Rotronics MP100H probes – four each, one spare.
o See Appendix G, Wiring Diagrams.
Data logger. The system uses a Campbell Scientific Instruments CR23X data logger.
Vista de página 2