
SECTION
14. INSTALLATION
AND MAINTENANCE
damage
the datalogger.
Campbell
Scientific's
DC112
phone
modem has spark
gaps
on the
phone
lines.
A 12
AWG wire should
be
run
from
the
modem
ground
terminalto earth
ground.
In
laboratory
applications, locating
a
stable
earth
ground
is not
always obvious. In
older
buildings,
new
cover
plates
on old
AC sockets
may indicate
that
a safety
ground
exists when
in
fact
the socket
is not
grounded.
lf
a safety
ground
does
exist, it is
good
practice
to verify
that it
carries no
current. lf the integrity
of the
AC
power
ground
cannot
be
verified,
it is better
to
ground
the system
to a massive metalobject
such
as a steelwater
pipe.
In the field,
an earth
ground
may
be created
through
a
grounding
rod. A
12
AWG
or
larger
wire
should be run
between the 21X
terminal
and
the eafth
ground.
Campbell
Scientific's
CM10
and CM6 Tripods
come complete with
ground
and lightning
rods,
grounding
wires,
and
appropriate
ground
wire
clamps.
14.8.2
EFFECT
OF GROUNDING ON
MEASUREMENTS:
COMMON
MODE RANGE
The common
mode range is the voltage
range,
relative
to the 21X
ground,
within
which
both
inputs
of
a differential measurement must
lie in
order for
the differential
measurement to be
made.
Common mode
range
for
the 21X is
IS.OV.
For example,
if the high side of
a
differential
input is
at 4V and the
low
side is
at
3V relative
to 21X
ground,
a
measurement
made
on the +5.0V range
would
indicate
a
signal of 1V. However,
if the
high
input
changed
to 6V,
the common
mode
range is
exceeded
and
the measurement cannot be
made.
Common
mode range
may be exceeded when
the 21X is
measuring
the output
from
a sensor
which
has its
own
grounded
power
supply
and
the low
side of the
signal is referenced
to
power
ground.
lf
the 21X
ground
and
the
sensor
ground
are at sufficiently
different
potentials,
the
signalwillexceed
the
common
mode range.
To
solve this
problem,
the sensor
power
ground
and
the 21X
ground
should
be
connected,
creating
one
ground
for
the system.
In
a laboratory
application,
where
more
than
one AC
socket
may be
used to
power
various
sensors,
it is not
always
safe to
assume
that the
14-6-
power
grounds
are at the same
potential.
To
safe, the
ground
of allthe AC sockets in use
should
be tied together with a 12 AWG wire.
14.9
MAINTENANCE
The 21X requires
a minimum of routine
rnaintenance.
A few
preventative
maintenance
steps willoptimize
battery life and decrease
the
chances
of datalogger failure.
The
21X alkaline
supply should
not drop
below
9.6V betore replacement.
When not in use,
remove
the eight
cells to eliminate
potential
corrosion of contact
points
and
store in a cool
dry
place.
When not in
use, the 21XL lead acid battery
should
be stored in a cool, dry environment with
the AC
charging circuit activated.
Discharging
the 21XL lead
acid batteries
below 11.76V
causes
permanent
damage,
or at least
drastically diminishes
the batteries'
normal life
expectancy.
When connecting
an external
battery to the
21X, use care
to avoid shorting
the +12V and
ground
from
the battery.
Four
packets
of active desiccant are
provided
with
each
21X.
Two
packets
are
located in
the
battery
base and two spare
packets
are sealed
in
a
plastic
bag. Used desiccant
may
be
reactivated
by oven drying for about 16 hours at
120oC. Routine
desiccant replacement is
required for reliable
datalogger service.
14.10
CALIBRATION
PROCEDURES
The 21X requires
little maintenance or
calibration. Measurements
are
made in such a
way
that small errors in the calibration are
automatically removed.
Measurements can be
made
to determine whether
the
accuracy of the
21X is within
the specifications
given
at the
of
this
manual.
lf needed,
the
calibration
procedures
described
in
this section
can
be
pedormed
by an experienced technician having
the suggested equipment.
NOTE: The
precision
of
the
21X
exceeds
that of most standard
electronic
equipment.
These
procedures
require that the
test
equipment have a
precision
and accuracy
equal to or better
than the 21X.
Comentários a estes Manuais