ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048 Page C-15
4.99 k
Settings for Temperature
ITEM DATA
COMMAND ENTERED
Item 1: Process Category Code 4 : Special Sensor
c1/4
Item 2: Process Type code = 1 : NTC Thermistor
c2/1
Item 3: Input for Thermistor Voltage = S1 : TEMP
1
c3/S1
1
Item 4: a-coefficient 0.001288
c4/0.001288
Item 5: b-coefficient 0.0002356
c5/0.0002356
Item 6: c-coefficient 9.557E-08
c6/9.557E-08
Item 7: R2 resistance 4990
c7/4990
1 This value will vary depending on the number of sensors being connected to the ZENO
-3200.
DATA OUTPUT MENU
Settings for Temperature
ITEM DATA
COMMAND ENTERED
Item 1: Field Type code = 9 : Transmit and Log Data Field
c1/9
Item 2: Output Message(s) = 1 DEFAULT
Item 3: Field Name = TEMP
c3/TEMP
Item 4: Data Process Record = P1.1
2
c4/P1.1
2
Item 5: Field Decimal Places = 1
c5/1
Item 6: Field Width = 5
c6/5
Item 7: Data Storage Class Code = 7 : Compressed Floating Point (2)
c7/7
DEFAULT = Default value, it is not necessary to enter a command for this item.
2 This value will vary depending on the number of sensors being connected to the ZENO
-3200.
AGND
CH2+
EXC+
Thermisto