ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page F-15
#12340001ENDSAMPLE60^C
The ZENO
®
-3200 would respond with:
#00011234OK,93^C
There may be a pause between when the ZENO
®
-3200 sends the ENDSAMPLE command and when the
CCSAIL
OK response is sent. This pause allows the ZENO
®
-3200 to complete any current tasks.
STARTSAMPLE,x,
This command instructs the remote ZENO
®
-3200 to check that its configuration is internally consistent,
and if so then to start sampling. This command is equivalent to leaving the Zeno Program Menu through
the standard User Interface.
The parameter
x takes one of the values Y (yes) or N (no).
The ZENO
®
-3200 will respond with a CCSAIL OK if the current ZENO
®
-3200 configuration is internally
consistent, and will immediately start sampling. This response is unaffected by the value of the parameter
(
Y or N). If the ZENO
®
-3200 is already sampling (i.e., the ENDSAMPLE was not previously used), it
responds with a CCSAIL
OK and continues sampling without pause.
If the current ZENOSOFT
®
configuration is not internally consistent, ZENOSOFT
®
returns with a
CCSAIL ERROR 14. If the value of the parameter was
N (no), ZENOSOFT
®
simply returns a standard
CCSAIL ERROR 14. If the value of the parameter was
Y (yes), ZENOSOFT
®
appends a set of warning
messages to the ERROR 14 (before the checksum). These are the same messages that appear if the user
attempts to quit the Zeno Program Menu with an inconsistent interface. In either case, the ZENO
®
-3200
does not re-start data sampling. The user must either correct the problems with the ZENO
®
-3200
configuration; or they must use the (pre-existing) CCSAIL
REBOOT command to reboot and reload the
previously saved configuration.
Two examples, with an inconsistent configuration, are as follows. If the user does not request diagnostic
messages, the following conversation occurs.
#06290001STARTSAMPLE,N,72^C
#00010629ERROR14,64^C
If the user does request diagnostic messages, the following conversation occurs.
#06290001STARTSAMPLE,Y,83
Data output record #20 Item 4, Sensor reference S3.2 is invalid.
#00010629